MUNICH, GERMANY, November 15th, 2024 / LARALAB GmbH, a leader in AI-powered medical imaging for interventional cardiology, announces the launch of its CE marked Aortic Valve Module. This module offers automatic, high-precision segmentations and measurements, crucial for advancing Transcatheter Aortic Valve Replacement in patients with Aortic Stenosis or Regurgitation | Read the press release
AI and Measurement Performance
Study: Fully automated CT analysis with a deep-learning based algorithm for pre-procedural planning in transcatheter aortic valve implantation [1]
98 patient dataset from University Hospital Frankfurt, all with symptomatic severe aortic stenosis undergoing TAVI.
Measurement | LARALAB (Mean±SD) | Expert (Mean±SD) | ICC |
---|---|---|---|
Mean Annulus Diameter (mm) | 24.4±2.4 | 24.0±2.4 | 0.97 |
Mean Annulus Perimeter (mm) | 77.7±7.4 | 76.1±7.5 | 0.96 |
Mean Annulus Area (mm^2) | 468.9±92.1 | 455.6±91.0 | 0.98 |